Nitric Oxide Sensor (NO Sensor) 4-Electrode - NO-B4

? Range: 20 ppm NO limit of performance warranty
? Sensitivity: 500 to 850 nA/ppm at 2ppm NO
? Response time t90: < 45 (s) from zero to 2ppm NO
? Temperature range: -30 to 40 ℃
? Pressure range: 80 to 120kPa
? Humidity range: 15 to 85% rh continuous
? Weight: < 13g

  • Model Number: NO-B4
  • Data Sheet: Date Sheet File
Quantity:
- +
  • Nitric Oxide Sensor (NO Sensor) 4-Electrode
  • Nitric Oxide Sensor (NO Sensor) 4-Electrode
Mouse over image to zoom
Share:

Product Specification

NO-B4 Nitric Oxide Sensor (NO Sensor) Description
Alphasense Nitric Oxide gas sensors operate using proven fuel cell technology.

Sensors are available in the following sizes:
A: 20mm diameter, the industry standard size for portable gas detectors
B: 32mm diameter package, the best choice for fixed site applications
D: miniature, with proven long-term performance. Sensors designed for the next generation of portable gas detectors.

The NO range provides OEMs with reliable sensors for use in a number of high volume applications. Strong signal levels combined with low zero current allows resolution to 200 parts per billion (ppb) whilst giving linear outputs to as high as 250ppm. The sensors are suitable for inclusion in fixed installation sensing heads, portable safety instruments, urban air monitoring and stack gas analysers.

Alphasense NO sensors provide a significant improvement in stability of response to transient relative humidity changes.

The NO sensors offer our electrolyte leak-free guarantee and reliable long-term detection performance. The leakproof housing is moulded with a colour coded top for ease of identification.

NO-B4 Nitric Oxide Sensor (NO Sensor) PERFORMANCE
Sensitivity nA/ppm at 2ppm NO 500 to 850
Response time t90 (s) from zero to 2ppm NO < 45
Zero current nA in zero air at 20℃ 30 to 140
Noise* ±2 standard deviations (ppb equivalent) 15
Range ppm NO limit of performance warranty 20
Linearity ppb error at full scale, linear at zero and 5ppm NO < ±1
Overgas limit maximum ppm for stable response to gas pulse 50
* Tested with Alphasense ISB low noise circuit



NO-B4 Nitric Oxide Sensor (NO Sensor) LIFETIME
Zero drift ppb equivalent change/year in lab air 0 to 50
Sensitivity drift % change/year in lab air, monthly test 0 to -20
Operating life months until 50% original signal  (24 month warranted) > 24



Nitric Oxide Sensor (NO Sensor) NO-B4  ENVIRONMENTAL
Sensitivity @ -20℃ (% output @ -20℃/output @ 20℃) @ 2ppm NO 60 to 90
Sensitivity @ 40℃ (% output @ 50℃/output @ 20℃) @ 2ppm NO 97 to 110
Zero @ -20℃ nA 0 to 30
Zero @ 40℃ nA 100 to 200



NO-B4 Nitric Oxide Sensor (NO Sensor) CROSS SENSITIVITY
H2S sensitivity % measured gas @ 5ppm H2S (after 3 minutes) < 10
NO2 sensitivity % measured gas @ 5ppm NO2 (after 3 minutes) < 4
Cl2   sensitivity % measured gas @ 5ppm Cl2 < 3
SO2 sensitivity % measured gas @ 5ppm SO2 < 5
H2   sensitivity % measured gas @ 100ppm H2 < 0.1
CO   sensitivity % measured gas @ 5ppm CO < 0.3
NH3 sensitivity % measured gas @ 5ppm NH3 < 0.1
CO2 sensitivity  % measured gas @ 5% Vol CO2   < 0.1
O3   sensitivity % measured gas @ 100ppb O3 < 4
Halothane sensitivity               @ 100ppm Halothane < 0.1



NO-B4 Nitric Oxide Sensor (NO Sensor) KEY SPECIFICATIONS
Bias voltage mV(working electrode potential is above reference electrode) 200
Temperature range -30 to 40
Pressure range kPa 80 to 120
Humidity range % rh continuous 15 to 85
Storage period months @ 3 to 20℃ (stored in sealed pot) 6
Load resistor Ω (ISB circuit is recommended) 33 to 100
Weight g < 13

Matched Transmitter board for NO-B4 NO Sensor

Individual Sensor Board (ISB) Alphasense B4 4-Electrode Gas Sensors
The Individual Sensor Board (ISB) is designed for use with the Alphasense B4 family of four-electrode gas sensors. This potentiostat provides a dual channel voltage output.
A low noise bandgap provides the bias voltage for NO sensors and the ISB can measure both oxidising (CO, H2S, SO2 and NO) and reducing (O3 and NO2) gases.
The ISB is configured for specific sensors: NO, NO2, O3 and CO / H2S / SO2 .
Designed for low power applications, the ISB requires 3.5 to 6.4 stable DC supply at only 1mA.


If you cannot find what you want, you can entrust ISweek to source for you. Just click:

Sourcing Service
国产精品成人久久久久| 一本大道无码日韩精品影视| 一区二区三区四区精品视频| 中文字幕精品一区二区三区视频| 国产精品午夜爆乳美女| 国产精品嫩草久久久久| 中日韩精品无码一区二区三区| 精品免费久久久久久久| 亚洲精品视频观看| 精品人妻va出轨中文字幕| 亚洲精品高清无码视频| 国产在线精品一区二区不卡麻豆| 亚洲精品第一国产综合精品99| 久久中文字幕2021精品| 国产精品女人呻吟在线观看| 久久国产免费观看精品3| 国产成人亚洲精品91专区高清| 国产99re在线观看只有精品| 国产精品香蕉在线观看| 国产精品电影一区| 99精品福利国产在线导航| 精品国产V无码大片在线看| 一本久久a久久精品亚洲| 久久精品国产WWW456C0M| 国产91在线精品| 色一乱一伦一图一区二区精品| 中文国产成人久久精品小说| 国产国拍亚洲精品福利| 2022精品天堂在线视频| 精品日韩在线视频一区二区三区| 精品人体无码一区二区三区| 亚洲精品无码乱码成人| 久久96国产精品| 一区二区精品在线| 中文字幕精品亚洲无线码一区| 毛片亚洲AV无码精品国产午夜| 精品国产品香蕉在线观看75| 曰韩精品无码一区二区三区| 亚洲国产精品一区二区第一页免| 国产精品va无码二区| 热99精品只有里视频最新|