HYDRA Ammonium Analyzer - HYDRA NH4-N

? NH4+-N? Measurement:? 0.01 to 14,000 ppm
? ?Potassium Measurement:? 0.01 to 40,000 ppm?
? pH Measurement:?0 to 14 pH?
? Temperature Measurement: 0°C to 100°C (32°F to 212°F)
? Operating Temperature: 0°C to 50°C (32°F to 122°F)
? Response Time: T90 1 minute

  • Model Number: HYDRA NH4-N
  • Data Sheet: Date Sheet File
Quantity:
- +
  • HYDRA Ammonium Analyzer
  • HYDRA Ammonium Analyzer
Mouse over image to zoom
Share:

Product Specification

HYDRA NH4-N Ammonium Analyzer Description
The HYDRA? Ammonium Analyzer measures the concentration of dissolved ammonium as nitrogen (NH4+-N) in water. The sensor uses three electrodes to determine the NH4+-N concentration, an Ammonium Ion Electrode, a Potassium Ion Electrode and a pH Electrode. It is designed for use in all kinds of water. Typical applications include monitoring environmental waters, lakes, streams and wells as well as wastewater treatment in aeration basins and effluent.

The Ammonium Ion Electrode provides the primary measurement. Any potassium ion in the sample generates a positive interference in the measurement, due to its similar size and charge to the ammonium ion. A Potassium Ion Electrode measures the amount of potassium ion present in the sample and HYDRA C22 Analyzer subtracts the appropriate amount of signal from the Ammonium Measurement.

HYDRA NH4-N Ammonium Analyzer Features
? Ammonium ISE electrode
? K+, pH and Temp. electrodes
? Rugged PVC design
? Integral Spray Head Cleaner
? Internal Signal Conditioning

HYDRA NH4-N Ammonium Analyzer Benefits
? Fast and Accurate Ammonium
? Measurement NH4+ or NH4+-N
? Fully compensated for pH, K+ interferences & Temperature
? Removable electrode guard for easy maintenance
? Clean sensor in situ with pressurized water or air
? Amplified signals allow up to 200 meters between Sensor and Analyzer

HYDRA NH4-N Ammonium Analyzer Specifications

Sensor
Sensor
Three Electrode system with spray cleaner, Ammonium ISE (NH4+-N) 
is the primary measurement. The Potassium ISE and pH glass electrodes 
are used to compensate the NH4+ signal. The Sensor is waterproof 
with an ingress rating of IP 68.
Measurement Range
NH4+-N: 0.1 to 14,000 ppm
Operating Temperature
0°C to 50°C (32°F to 122°F)
Min/Max Flow Rate
Minimum 0.1 m/s
Maximum 3.0 m/s
We&ed Materials
PVC, PES, PVDF, PTFE, Viton, Glass, 316 SS
Accuracy
± 3% of reading, dependent on Calibration
Response Time
T90 1 minute
Electrode Life
ISEs: 4- 6 months, typical
pH electrode: 6-12 months, typical
C22 Analyzer
Measurements

    Ammonium:
0.01 to 14,000 ppm as NH4+-N
    Potassium:
0.01 to 40,000 ppm
    pH:
0 to 14 pH
    Temperature:
0°C to 100°C (32°F to 212°F)
Compensation

    pH:
4 - 10 pH
    Potassium:
0.1 to 1000 ppm
Display
2.5" X 1.75" backlit LCD, 4 lines of Text & Graphical
Enclosure
NEMA 4X, L x W x D: 5.7" x 5.7" x 7
Outputs
(2) 4-20 mA maximum load 800 ohms @ 24 VDC
Configured: 0.1 to 20 mg/l NH4+-N; 0 - 14 pH
Optionally up to (4) 4-20 mA outputs*
Input Power
110/220 VAC @ 50/60 Hz
Alarm Relay Ra%ngs
(2) SPDT 230 VAC/5A or 30 VDC/5A resistive max.
Relay(1) Spray Cleaner, Relay(2) Alarm
Optionally up to (8) Relays*

Part No.
Model and Product Description
1290030-1
HYDRA? NH4-N Sensor, complete, NH4, K+, pH, Temp, Spray Cleaner head and 30 #. cable
1290030-2
HYDRA? NH4-N Sensor, complete, NH4, pH, Temp, Spray Cleaner head and 30 #. cable (No K+ Sensor)
16KA2221.K100
HYDRA? C22 NH4-N Analyzer, K+ compensated, (2) 4-20 mA output, 0.1 - 20 ppm NH4-N and (2) relays*
16KA2221.J100
HYDRA? C22 NH4-N Analyzer, (2) 4-20 mA output, 0.1 - 20 ppm NH4-N and (2) relays* (No K+ compensation)

Part No.
Spare Parts and Accessories Description
2005083.VIT
Ammonium Electrode Cartridge (recommended spare)
2005034.VIT
Potassium Electrode Cartridge (recommended spare)
2005145.VIT
pH Electrode Cartridge (recommended spare)
3300854-1
Replacement Spray Nozzle
3501050-1
PVC Front Sensor Guard
2010445-1
Ammonium Calibration solution, NH4-N 1 ppm
2010446-1
Ammonium Calibration solution, NH4-N 100 ppm
2010443-1
Potassium Calibration solution, 1 ppm
2010444-1
Potassium Calibration solution, 100 ppm
2010100
pH 4 Buffer Calibration solution
2010101
pH 7 Buffer Calibration solution
1000300-1
4-20 mA USB Data Logger

If you cannot find what you want, you can entrust ISweek to source for you. Just click:

Sourcing Service
久久精品一区二区三区四区 | 国产视频精品免费视频| 国产精品大全国产精品| 在线精品视频播放| 久久国产精品99久久小说| 日韩精品无码一区二区三区AV| 日本加勒比久久精品| 国产真实乱子伦精品视 | 亚洲性色精品一区二区在线| 亚洲国产精品成人久久久| 99re5精品视频在线观看| 久久国产视频精品| 久久精品视频大全| 久久久精品人妻一区亚美研究所| 精品国产自在在线在线观看| 国产福利精品一区二区| 久久精品国产一区二区三| 无码国内精品久久人妻| 99国产精品视频免费观看| 99re视频精品全部免费| 在线人成精品免费视频| 亚洲午夜久久久精品电影院| 四虎精品视频在线永久免费观看| 国产成人精品综合网站| 伊人久久精品无码麻豆一区| 精品久久久久久久无码久中文字幕 | 久久97久久97精品免视看秋霞| 国产成人精品久久| 久久亚洲AV永久无码精品| 国产亚洲美女精品久久久2020| 国产精品揄拍100视频| 国产精品9999久久久久| 精品性影院一区二区三区内射| 91精品国产91久久久久久青草| 精品国产一区二区三区无码| 一色屋成人免费精品网站| 精品久久久久久无码人妻中文字幕 | 中文人妻熟妇乱又伦精品| 亚洲精品色在线网站| 精品免费AV一区二区三区| 国产精品亚洲高清一区二区|